Key features

  • Pressure Source and mA Measurement to calibrate pressure devices
  • Built-in pump for sourcing pressure to 300 PSI
  • Easy-to-clean, field-serviceable, integrated pressure pump
  • Built-in switch test for calibrating pressure switches
  • Extend pressure capabilities by using Fluke 750P pressure modules

Product overview: Fluke 718 Pressure Calibrator

The complete solution to pressure calibration

The Fluke 718 Pressure Calibrator provides a total pressure calibration solution for transmitters, gauges, and switches. This compact pressure calibration solution is about a third of the size of comparable instruments and weighs just one kilogram (two pounds). The unique easy-to-clean pump helps protect it from damage and provides the ability to service the pump without disassembly, even in the field.

With an easy-to-use error % calculation you can make pass/fail decisions quickly. Meaning you can make decisions in the field without having to analyze the results later. The min/max/hold function also makes it easy to capture changing measurements for comparison or documentation. The integrated pressure switch test function, and best-in-class mA accuracy (0.015%) provide a high level of measurement confidence and even broader workload coverage.

Extend your pressure capabilities

The Fluke 718 Pressure Calibrator is available in 1, 30, 100, and 300 PSI ranges. It is also compatible with Fluke 750P Series Pressure Modules so you are certain to find a model that will meet your calibration needs.

Related resources

What's in the box:

  • Fluke 718 Pressure Calibrator
  • Protective Boot (installed)
  • (2) 9V batteries (installed)
  • Traceable calibration certificate w/data
  • (1 pair) TL75 Test Leads
  • (1 set) AC175 Alligator clips
  • 3-year warranty
  • Quick Reference Guide

Specifications: Fluke 718 Pressure Calibrator

Technical Specifications
Fluke 718 Series
Measure pressure (Internal sensor)
Accuracy 0.025% full scale
Functions Zero, min/max, hold, damp
Dry air, non-corrosives only
Fluke 718 1G
Range -1 psi to +1 psi, -68.9 mbar to 68.9 mbar (-6.89 kPa to 6.89 kPa)
Resolution 0.0001 psi, 0.001 mbar, 0.0001 kPa
Over pressure 5x Full Scale
Functions Zero, Min, Max, Hold, Damp
Fluke 718 30G
Range -12 psi to 30 psi, (-850 mbar to 2 bar, -85 to 206.84 kPa)
Resolution 0.001 psi, 0.1 mbar, 0.01 kPa
Over pressure 5x Full Scale
Functions Zero, Min, Max, Hold, Damp
Fluke 718 100G
Range -12 psi to 100 psi, (-850 mbar to 7 bar, -85 to 689.48 kPa)
Resolution 0.01 psi, 0.1 mbar, 0.01 kPa
Over pressure 2x Full Scale
Functions Zero, Min, Max, Hold, Damp
Fluke 718 300G
Range -12 psi to 300 psi, (-850 mbar to 20 bar, -85 to 2068.42 kPa)
Resolution 0.01 psi, 1 mbar, 0.1 kPa
Over pressure 375 psi 25 bar
Functions Zero, Min, Max, Hold, Damp
Fluke 718 series via 700 Series Pressure Modules
Range 29 pressure modules, 0 – 1 inH20 to 10,000 psi, 2.5 mbar to 700 bar. Over pressure per pressure module specs
Resolution Per pressure module specs
Accuracy To 0.025% of full span, per pressure module specs
Functions Zero, Min, Max, Hold, Damp
Media compatibility per pressure module specs
Fluke 718 Series Via Built in Pump
Range -12 psi or -850 mbar to full scale Supported Pressure Units; psi, inH20 (4°C), inH20 (20°C), cmH20 (4°C), cmH20 (20°C), bar, mbar, kPa, inHg, mmHg, kg/cm
Loop power range 24 V DC
Accuracy ±10%
Drive; 20 mA into 1000 Ω for battery > 6.8 V; 700 Ω for battery 5.8 to 6.8 V
mA Measure
Range 0 mA to 24 mA
Resolution 0.001 mA
Accuracy 0.015% + 1 count
General Specifications
Maximum voltage 30 volts (non-operating)
Temperature -40°C to 60°C
Operating temperature -10°C to 55°C
Relative humidity 95% (10°C to 30°C); 75% (30°C to 40°C); 45% (40°C to 50°C); 35% (50°C to 55°C)
Operating altitude 3,000 M max
Shock 1 M drop test
Vibration Random, 2 g, 5 Hz – 500 Hz
Safety CSA C22.2 No. 1010.1: 1992
EMC EN50082-1: 1992 and EN55022: 1994 Class B
Size 216 x 94 x 66 mm (8.50 x 3.72 x 2.60 in)
Weight 992 g (35 oz)
Power, battery life 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718
Battery life 4 to 20 hours typical, depending on functions used
Warranty Three years (one year on pressure pump in Fluke 718)
Display LCD, 5 digit pressure and current simultaneous
For use with non-corrosive gasses

Models: Fluke 718 Pressure Calibrator

Fluke 718 1G
Fluke 718 1G Pressure Calibrator
-1 to 1 psi, (-68.9 mbar to 68.9 mbar, -6.89 to 6.89 kPa)

Includes:

  • Protective Yellow Holster with test lead storage
  • Instructions Sheet (14 languages)
  • Test Leads and Alligator Clip
  • Two 9V alkaline batteries
  • NIST - Traceable calibration report and data
Fluke 718 30G
Fluke 718 30G Pressure Calibrator
-12 psi to 30 psi, (-850 mbar to 2 bar, -85 to 206.84 kPa)

Includes:

  • Protective Yellow Holster with test lead storage
  • Instructions Sheet (14 languages)
  • Test Leads and Alligator Clip
  • Two 9V alkaline batteries
  • NIST - Traceable calibration report and data
Fluke 718 100G
Fluke 718 100G Pressure Calibrator
-12 psi to 100 psi, (-850 mbar to 7 bar, -85 to 689.48 kPa)

Includes:

  • Protective Yellow Holster with test lead storage
  • Instructions Sheet (14 languages)
  • Test Leads and Alligator Clip
  • Two 9V alkaline batteries
  • NIST - Traceable calibration report and data
Fluke 718 300G
Fluke 718 300G Pressure Calibrator
-12 psi to 300 psi, (-850 mbar to 20 bar, -85 to 2068.42 kPa )

Includes:

  • Protective Yellow Holster with test lead storage
  • Instructions Sheet (14 languages)
  • Test Leads and Alligator Clip
  • Two 9V alkaline batteries
  • NIST - Traceable calibration report and data
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 29.42ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, , 29.44ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned; Content Type "questions" is not supported by getReviews().;
Please enable cookies to view product reviews.